33
REPLACING DELIVERY REGULATOR (REPAIR)
If it is necessary to change the delivery pressure regulator due to a regulator fault, as indicated on the front of the Medical
IntelliSwich, perform the following steps:
CAUTION: Resetting the delivery regulators should be performed after replacing and while the Medical IntelliSwitch is flowing gas.
Adjustment in a static condition (not flowing) will effect the available pressure during max flow conditions.
WARNING: Replacement of faulted regulator should be done as soon as possible to insure uninterrupted gas supply. If the second
regulator should fail while the first is out for repair, loss of line pressure and/or line over-pressurization will occur.
1.
Open door by loosening the four screws holding it to the cabinet.
2.
Locate delivery regulator toward the top of the system, see figure 25. Isolate the faulted regulator by closing the inlet
and outlet quarter turn valves for that side.
Figure 25
Содержание IntelliSwitch 57 Series
Страница 5: ...5 Figure 1 ...
Страница 25: ...25 Figure 14 ...
Страница 30: ...30 Figure 20 ...
Страница 37: ...37 INTERNAL CABLE ASSEMBLIES REPAIR Figure 31 ...
Страница 38: ...38 Figure 32 Figure 33 ...
Страница 50: ...50 SELF TEST WORKSHEET Page 1 ...
Страница 51: ...51 SELF TEST WORKSHEET Page 2 ...