27
CHANGING AC INPUT POWER (REPAIR)
The Medical IntelliSwitch™ has the AC Supply Voltage factory set based on the product part number.
CAUTION!
Setting the AC Supply voltage incorrectly can cause permanent damage to the product. If changing the supply voltage
make sure the supply switch correctly matches the AC voltage being supplied to the product.
In the event that the AC Supply Voltage needs to be changed, perform the following steps.
1.
Turn AC power OFF to the system.
2.
Open front door by loosening the 4 screws holding it to the cabinet.
3.
Locate the AC power switch on the Control Board (Figure 16).
4.
Move the switch to the proper setting. THIS SWITCH MUST BE SET TO THE AC INPUT VOLTAGE BEING SUPPLIED
TO THE PRODUCT.
5.
Make sure the system is correctly wired to the AC Supply Voltage.
6.
Close front door and re-secure it to the cabinet.
7.
Re-apply AC power to the system.