11.
The display will now show the actual measured value with unit.
In order to avoid unrequested alarms make sure the signal created by the test gas has
dropped below the alarm threshold prior to returning into the measuring mode.
12.
Confirm with
to return to the main menu.
13.
Return into the measuring mode by pushing Reset.
If the calibration has not been completed successfully (ERROR 8, CALIBR. FAILED or GAS
CONC ERROR), the module will continue with the most recent valid data. The display shows
alternating „CALIB. FAILED“ and the measured value. A new calibration should be done as
soon as possible.
Reading the bridge voltage
Active in the programs for Statox 501 HRC, ARE, LCIR, MCIR, PID sensor heads.
The current bridge voltage of the sensor can be displayed.
Figure 14: Bridge voltage menu diagram
BRIDGE
VOLTAGE
(-) xxx
mV
R
R
+
!
!
MEASURING MODE
Change Menu item
Actual bridge voltage,
no Time-Out