Statox 502 Control Module
Page 28 of 40
Issue 1/2015
Current output in Service Mode
You can select which current the analog output generates in the service mode. The ex – works setting is
2 mA. The selected output current is conserved independent from selected program or operation mode.
If you operate the module as a SIL device EN 50402 or IEC 61508, 2 mA in service mode are
mandatory.
Figure 20: Diagram service output
1.
Push
to enter the „SERVICE OUTPUT“ menu.
2.
The actual setting is shown on the display.
3.
Select a new setting with
◄
-
or
►
.
4.
Confirm with Enter.
5.
Push reset to return to the measuring mode.
!
MEASURING MODE
Change Menu item
Selection of current value