Statox 502 Control Module
Page 18 of 40
Issue 1/2015
Calibration
In this menu you can perform a calibration of sensor heads type Statox 501 HRC, ARE, LCIR, MCIR and PID
detecting combustible gases, carbon dioxide (CO
2
) and volatile organic compounds (VOC). The test gas
concentration must be within the permitted range, see program listing.
Figure 13: Calibration menu diagram
CALIB.
ROUTINE
ZERO
ADJUSTM.
000
APPLY
SPAN GAS
GAS
R
R
CALIB.
FINISHED
GAS CONC
(value) (unit)
FACTOR
x.yy
DONE
R
CALIBR.
FAILED
GAS
(value) (unit)
R
R
GAS CONC
ERROR
R
+
Change Menu item
MEASURING MODE
Flashing, searching zero point
Flashing, turn test gas on
Displayed if zero point is adjusted
Signal is stable,
Turn test gas off
Input: gas concentration (value)
Start zeroing
Input: Reference factor (value)
Correction
Confirmation
Displayed for 2 s after
successful calibration
Actual measuring value and unit