IE-MLG-M1200E-01
45
Mechanical characteristics
Dimensions (h x w x d):
178 x 95 x 37 mm without cables and connectors
Weight:
approximately 330 g, including batteries
Protection:
IP20
Included mounting accessories:
MLP001 - wall holder for Multilogger
Optionally mounting accessories:
MLP002 - lockable wall holder for Multilogger
MLP003 - magnetic holder for Multilogger
Technical parameters of the inputs
The device is equipped with a white, uncompensated miniature
thermocouple connector. Compensation of the cold junction temperature is
performed by measuring the temperature inside of the device in the area
between the input connectors. The inputs are not galvanically isolated from
each other, nor from other signals connected to the device. In order to secure
correct measurements, you need to prevent undesirable coupling.
Furthermore, the Multilogger has to be placed in its working position and
temperature stabilized.
Required measured quantities can be selected for each input using the
COMET Vision
software. These quantities can be recalculated using two-
point linear transformation, displayed, recorded to the memory or used for
alarming. In addition, the Multilogger allows creating of calculated quantities
from the measured values using simple mathematical operations.
The Multilogger allows measure
temperature in °C or in °F. If you use °F, the
below stated ranges and accuracy have to be recalculated to
°F.