18
Treadmill Hardware Manual 1011-0002 v1.02
Treadmill Stimulus Detection with AC power cord
1 x CAT5 network cable, 0.3m (1ft), 25-pin
1 or more detection cables, 1.8m (6ft), 25-pin
1 x Instruction Manual
To connect the Treadmill Stimulus Detection to the Treadmill Controller:
•
Switch OFF the Treadmill Controller and remove all the shocker cables from the treadmill system.
•
Place the Treadmill Stimulus Detection on top of the Treadmill Controller. Be sure to center the
cabinets feet over the top of the locating pins.
•
Using one of the shocker
cables, connect the “Shocker” port of the Treadmill Controller to the “From
Controller” port of the Treadmill Stimulus Detection.
(other cables removed for clarity)
•
Remove the CI-Bus terminator from the Treadmill Controller.
•
Use the short 0.3m (1ft) CAT-5 network cable and connect
the adjacent “CI
-
Bus” ports of the Treadmill
Stimulus Detection and Treadmill Controller together.
•
Plug the CI-Bus terminator into the top port of the Treadmill Stimulus Detection.
Содержание Exer 3/6
Страница 2: ......