Page: 84
EnergyTech 401
CODEL
Operating and Maintenance Manual
Monitoring Solutions
Set-Up Mode
5.10
All operating parameters – averaging times, output settings, normalisation
parameters, path length, calibration, etc. – can be changed from this mode. To
prevent any unauthorised changes, the user must enter a four number code
before the mode can be entered.
Figure 19 : Set Up Mode Tree
10 to 60s @ 10s i ntervals
01 to 60m @ 01m intervals
01 to 24h @ 01h i ntervals
01 to 30d @ 01d intervals
0 or 4-20mA
Take average from secs, mins, hours or days stack
ppm - mg/m3 - mg/Nm3
0 to 9999 span (ppm or mg/m3 - depends on above)
Measured, zero mA, FS (20mA) or fix value
Set up current output levels
4-digit code
01 to 30 (unique number for serial data, I/Ps & O/Ps)
Gas path length in mm
Alarm from secs, mins, hours or days stack
0 to 9999 (ppm or mg/m3 - depends on units)
Determines basic calibration of the instrument
Contact, serial, temp., O2 or Det. Level
Model 1030 only
Set up normalisation I/P & std. levels for temp.
Set up normalisation I/P & std. levels for O2
Set up normalisation I/P & std. levels for pressure
Set up normalisation I/P & std. levels for H2O
Reset the four stack averages
Detector D2
Adjust sensitivity
Model 1021 only
Conduct a basic calibration
Cali brate H2O (model 1030 only)
Seconds
Minutes
Hours
Days
Output
Average
Units
Span
Fault Condi tion
Set mA O/P
Security No.
Identity No.
Path Length
Alarm Source
Level
Cal. Factor
Plant Status
Display NO/NOx
Temperature
Oxygen
Pressure
Water Vapour
Yes
No
Set Detectors
Span Adjust
Auto Zero
Cali brate
Set Averages
Configure O/P
Parameters
Normali sation
Reset Average
Cali brate
Exit options not shown
Enter
Press Enter
on Exit option
After this mode has been selected, the instrument will suspend its operation and the Data Valid LED will extinguish. If
no key is pressed within 5 seconds after selection of this mode, the instrument will revert to the normal operating
mode.
NOTE
Содержание Energy Tech 401
Страница 71: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 71 ...
Страница 99: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 99 ...
Страница 101: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 101 ...
Страница 115: ...Section 6 Maintenance Issue A Revision 1 Date 24 07 19 Ref 100606 Page 115 Corrective Maintenance ...
Страница 118: ...Page 118 EnergyTech 401 CODEL Operating and Maintenance Manual Monitoring Solutions ...