Page: 28
EnergyTech 401
CODEL
Operating and Maintenance Manual
Monitoring Solutions
Pressure, temperature, and oxygen values can be held to normalise the
calculated gas value to standard conditions. These values may be read by the
instrument using the following methods:
fixed value from the keypad
4-20mA outputs from measurement transducers - the ranges represented by
these inputs are set from within the processor - inputs are taken to the signal
processor
if the analyser is part of an integrated system, the serial data line can carry the
normalising values
The plant status input facility is available to prevent the rolling average stacks
being diluted by measurements made during periods where the plant is shut
down. It is governed by one of three choices; a serial input (from an integrated
system), the logic input (terminals PS1 and PS2 in the signal processor) and
multiple.
Multiple has four options; temperature, oxygen, water vapour thresholds and
logic input. It is set in Mode 5.
In normal operation (plant operating), the plant status will register as ON.
However, if the plant status condition is broken, the status will change of OFF and
the averaging stacks (minutes, hours, days) will not be updated.
3.2.3
Normalising Inputs
3.2.4
Plant Status Input
For normal operation terminals PS1 and PS2 must not be linked together.
WARNING!!!
Содержание Energy Tech 401
Страница 71: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 71 ...
Страница 99: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 99 ...
Страница 101: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 101 ...
Страница 115: ...Section 6 Maintenance Issue A Revision 1 Date 24 07 19 Ref 100606 Page 115 Corrective Maintenance ...
Страница 118: ...Page 118 EnergyTech 401 CODEL Operating and Maintenance Manual Monitoring Solutions ...