Section 5
Normal Operation
Issue: A Revision: 1 Date: 24/07/19 Ref: 100606
Page: 91
Select the units for the alarm – these may be different to the units selected for
the analogue output.
After the source has been selected, the instrument requires a level that the
output will be triggered. Set the desired level with the ARROW keys.
During the calibration routine, the instrument calculates a ‘Cal Factor’ which sets
the basic calibration of the instrument. This value may be changed from this
mode.
Used to determine whether the plant is operating under correct conditions. If it is
not, plant status will be OFF and the minutes, hours and days averaging stacks
will not be updated. If data is stored whilst the plant status is OFF diluted overall
readings will be registered.
5.10.15
Cal Factor
5.10.16
Plant Status Input
Do not enter this mode without first recording the original cal factor – this value is displayed in Mode 4 – see
Calibration Data. The Cal factor will be lost when this menu option is entered.
As this value controls the calibration of the instrument do not change it without due consideration.
WARNING!!!
Содержание Energy Tech 401
Страница 71: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 71 ...
Страница 99: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 99 ...
Страница 101: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 101 ...
Страница 115: ...Section 6 Maintenance Issue A Revision 1 Date 24 07 19 Ref 100606 Page 115 Corrective Maintenance ...
Страница 118: ...Page 118 EnergyTech 401 CODEL Operating and Maintenance Manual Monitoring Solutions ...