Section 5
Normal Operation
Issue: A Revision: 1 Date: 24/07/19 Ref: 100606
Page: 69
5
Normal Operation
Introduction
5.1
After the instrument has been commissioned, it will measure the gas levels
between the transmitter and receiver and produce an output proportional to the
gas levels. An integral 32-character display also shows the calculated levels.
The instrument allows the operator to interrogate the micro-processor to
observe the system parameters and to change them if required.
A menu-based program is used and access is gained by four keys mounted on
the lid of the signal processor.
Once commissioning has been concluded, the absorption of IR radiation is
measured and a parameter ‘Y’ is calculated – see Principles / Modes of
Operation. This value is used to produce a final concentration of gas that can be
normalised to standard conditions and averaged over a time ranging from 10
seconds to 30 days.
The instrument computes four averages any of which can be used to drive the
analogue output, or displayed on the integral 32-character LCD.
During the commissioning procedure a calibration is conducted that sets the
system gains to produce a zero or known gas level. Once the routine has been
conducted the calibration of the instrument is fixed by precision filters, which do
not change.
5.1.1
Measurement
5.1.2
Calibration
Содержание Energy Tech 401
Страница 71: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 71 ...
Страница 99: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 99 ...
Страница 101: ...Section 5 Normal Operation Issue A Revision 1 Date 24 07 19 Ref 100606 Page 101 ...
Страница 115: ...Section 6 Maintenance Issue A Revision 1 Date 24 07 19 Ref 100606 Page 115 Corrective Maintenance ...
Страница 118: ...Page 118 EnergyTech 401 CODEL Operating and Maintenance Manual Monitoring Solutions ...