CME NIKI T34 Скачать руководство пользователя страница 6

 
 
 

C – Precautions

 

Information to ensure safe and effective use. 
 

 

Before use, the user must become thoroughly familiar with the operating instructions contained 
in the T34 user manual. 

 

The provider is responsible for ensuring that the T34 pump and medications are used according 
to the physician’s infusion therapy prescription. 

 

The provider is responsible for educating the patient in proper use of the pump. 

 

Do not allow water or other fluids to get inside the pump or the lockbox. Fluids getting into the 
pump may damage it. If the pump gets wet, stop the pump and initiate service of the pump. 

 

A security code must be keyed in to access the lock-level 2 programming functions. This code 
is  designed  to  allow  access  to  the  provider  only.  To  prevent  a  patient  from  tampering  with 
programmed settings, this security code should not be given to the patient. 

 

As  with  any  infusion,  the  disposables  used  with  the  T34  pump  must  be  compatible  with  the 
medication  being  delivered.  Check  with  the  manufacturer  of  the  disposables  before  use. 
Consult  the  drug  manufacturer’s  insert  for  precautions,  guidelines,  and  instructions  for 
preparing and using the medication. 

 

Do not drop the pump, strike it against hard objects or otherwise subject it to extreme physical 
stress. This could damage the pump, which may  impair the operation of the pump and/or the 
accuracy of the infusion.  If the pump is dropped, return the pump for inspection by qualified 
service personnel. 

 

Unsafe  operation  may  result  from  using  improper  accessories.  Use  only  accessories  and 
options designed for this system and supplied by the T34 distributor. 

 

A  kinked/occluded  infusion  line  may  impair  the  operation  and  accuracy  of  the  pump.  Before 
operation, verify that the infusion line is not kinked or occluded. 

 

Make sure any adjustments, maintenance or repair of the device are carried out by authorized 
service technicians certified by the T34 distributor. Adjustments, maintenance or repair made 
by un-qualified service personnel may impair the operation of the pump and/or the accuracy of 
the infusion. 

 

Battery damage could occur at ambient temperatures >55°C (131°F). 

 

To prevent damage to the pump, do not store the pump for extended periods with the battery in 
the pump. 

 

The  pump  housing  is  not  hermetically  sealed  –  immersing  the  pump  in  liquid  could  cause 
damage internal components. Do not immerse the pump into any type of liquid. 

 

If fluid is spilled on the pump, clean and dry it immediately. A damp cloth should be used for 
cleaning and a mild detergent may be used if needed. Xylene, acetone or similar solvents could 
damage the pump – do not clean the pump with harsh chemicals. 

Содержание NIKI T34

Страница 1: ...NIKI T34 SYRINGE PUMP Instruction Manual Ref 100 090SS Edition June 2008 0473...

Страница 2: ...Batteries 10 I Access Codes Keypad Lock 11 J Infusion set up programming 12 K Using Accessories with T34 Syringe Pump 16 L Recommended Checks during infusion 17 M Advanced Functions of the T34 18 N Ch...

Страница 3: ...adults or for pediatric use The T34 syringe pump can be easily concealed beneath clothing in a shoulder holster and combination belt A compact lockbox is provided to protect the Pump from damage and o...

Страница 4: ...that unintended programming and key presses do not occur The T34 syringe is contained within a locked box Only the healthcare provider is able to open the box After the healthcare provider programs th...

Страница 5: ...nt may cause serious injury or death Before connecting to the patient purge all air from the syringe and the administration set Depending on the route of administration and the volume of air air embol...

Страница 6: ...e subject it to extreme physical stress This could damage the pump which may impair the operation of the pump and or the accuracy of the infusion If the pump is dropped return the pump for inspection...

Страница 7: ...3 ON OFF key Switches the pump on and off 4 NO STOP key a Takes user back a step b stops pump 5 YES START key Confirms choices 6 Down arrow key scrolls between options 7 Up arrow key Scrolls between...

Страница 8: ...ttery Do not dispose of battery in municipal waste Symbol indicates separate collection for battery is required The use of single use disposable components on more than one patient is a biological haz...

Страница 9: ...Precautions Administration Sets Administration sets must be compatible with the medicine delivered Please check with the manufacturer of the infusion set before use Visually inspect packaging and cont...

Страница 10: ...y and need to be replaced more often Do not use non alkaline or rechargeable batteries There is some variation in size between different brands of batteries Never try to force a battery into the batte...

Страница 11: ...nu used for service and or repair of the Pump is also protected by a special access procedure and different code from Change Set Up Keypad Lock The T34 allows all users to lock the operation of the ke...

Страница 12: ...tus 3 Load the syringe into the Pump prior to connecting the syringe to the patient 4 The LCD display will show PRE LOADING and the actuator will start to move Wait until it stops moving and the syrin...

Страница 13: ...he programmed infusion rate the user can purge the system once only up to a volume limit configured under Change Set Up max 0 5 ml To use this option after syringe confirmation a Press FF key b Ensure...

Страница 14: ...l show the default 16 Use the UP DOWN arrow keys to set the desired duration or press YES to confirm acceptance of the default duration Setting the Infusion Rate 17 The Pump calculates and displays th...

Страница 15: ...START to commence infusion 23 While running the LCD displays infusion Time Remaining top line Infusion Rate main central line and the bottom line will alternate between Syringe Size Brand and Pump De...

Страница 16: ...the T34 place it into the lockbox and carry pouch to protect the syringe Pump syringe contents from damage or tampering 1 Open the lockbox using the standard key that operates all T34 lockboxes 2 Plac...

Страница 17: ...ing the INFO key either a third press or waiting a few seconds returns the display to the main infusion display screen Excessive key presses or usage of the INFO feature will reduce battery life Use o...

Страница 18: ...programming to set a max min rate To titrate the rate up or down during infusion follow these steps Use the UP or DOWN arrows on the keypad to adjust the rate to the desired new setting as displayed...

Страница 19: ...es all events in log are correctly date and time stamped ensure local update for DST FF Key Operation Limits the forward movement of actuator caused by pressing the FF key during syringe loading 0 1 1...

Страница 20: ...Turn pump off Syringe Displaced Check Syringe Press YES to confirm One or more of the syringe detection sensors is not detecting up the syringe Check the syringe and re seat as necessary Pump Paused T...

Страница 21: ...new battery Service required The infusion is going too quickly has ended early or too slowly volume remaining in syringe at end of infusion Incorrect rate set Wrong syringe brand onfirmed during set...

Страница 22: ...ry Pump Paused Too Long Syringe Displaced Syringe Empty T34 Dimensions 169 x 53 x 23mm Classification Type CF Equipment degree of protection against electrical shock Housing ABS fire retardant Weight...

Страница 23: ...he patient and or the patient s assistants or family members in case of questions or concerns Training the Patient and or the Patient s Assistants or Family Members 1 Using the Patient Guide instruct...

Страница 24: ...ith a small dry brush to remove debris or other particles Between Patients Change the battery Before connecting the pump to a patient and periodically during use clean the unit using a lint free cloth...

Страница 25: ...returned to CME or its designated representative should be properly packaged using CME shipping cartons and inserts Inadequate packaging may result in damage to the pump This Limited Warranty shall no...

Страница 26: ...trained certified service centers or biomedical engineers Please consult your local certified engineers or return to your T34 distributor should this need arise For alternate site homecare application...

Отзывы: