C – Precautions
Information to ensure safe and effective use.
Before use, the user must become thoroughly familiar with the operating instructions contained
in the T34 user manual.
The provider is responsible for ensuring that the T34 pump and medications are used according
to the physician’s infusion therapy prescription.
The provider is responsible for educating the patient in proper use of the pump.
Do not allow water or other fluids to get inside the pump or the lockbox. Fluids getting into the
pump may damage it. If the pump gets wet, stop the pump and initiate service of the pump.
A security code must be keyed in to access the lock-level 2 programming functions. This code
is designed to allow access to the provider only. To prevent a patient from tampering with
programmed settings, this security code should not be given to the patient.
As with any infusion, the disposables used with the T34 pump must be compatible with the
medication being delivered. Check with the manufacturer of the disposables before use.
Consult the drug manufacturer’s insert for precautions, guidelines, and instructions for
preparing and using the medication.
Do not drop the pump, strike it against hard objects or otherwise subject it to extreme physical
stress. This could damage the pump, which may impair the operation of the pump and/or the
accuracy of the infusion. If the pump is dropped, return the pump for inspection by qualified
service personnel.
Unsafe operation may result from using improper accessories. Use only accessories and
options designed for this system and supplied by the T34 distributor.
A kinked/occluded infusion line may impair the operation and accuracy of the pump. Before
operation, verify that the infusion line is not kinked or occluded.
Make sure any adjustments, maintenance or repair of the device are carried out by authorized
service technicians certified by the T34 distributor. Adjustments, maintenance or repair made
by un-qualified service personnel may impair the operation of the pump and/or the accuracy of
the infusion.
Battery damage could occur at ambient temperatures >55°C (131°F).
To prevent damage to the pump, do not store the pump for extended periods with the battery in
the pump.
The pump housing is not hermetically sealed – immersing the pump in liquid could cause
damage internal components. Do not immerse the pump into any type of liquid.
If fluid is spilled on the pump, clean and dry it immediately. A damp cloth should be used for
cleaning and a mild detergent may be used if needed. Xylene, acetone or similar solvents could
damage the pump – do not clean the pump with harsh chemicals.