10ml BD Plastipak

Duration   24:00

 Change ↑↓   Press YES 

10ml BD Plastipak

Rate           0ml/h

 Change ↑↓   Press YES

 

Duration Changed

 

Please Check rate

 

Volume             12 ml
Duration            24:00   
Rate                    0.5 ml/h

   

Confirm,

 

Press YES

 

WARNING

: If the volume displayed after loading the syringe is greater than or 

significantly less than the volume visually confirmed on the syringe scale, remove the 
syringe, turn off the pump and, with the barrel clamp arm down, turn the pump on to 
allow  pre-loading  to  occur.  Repeat  the  Syringe  placement  and  detection  steps  and 
ensure  the  correct  syringe  size  and  brand  are  confirmed.  If  the  calculated  volume 
reading  is  still  significantly  different  from  the  visually  confirmed  contents,  remove 
the pump from use and return to an authorized service centre for inspection, testing 
and calibration. 

 
14.   Press YES/START key to confirm the volume to be infused (VTBI). 
 
15.   The  LCD  display  now  prompts  the  healthcare  provider  to  set 

the  Duration  of  the  infusion.  The  Pump  is  likely  to  be 
configured to a default duration in the access code protected set 
up  menu.  For  example,  senior  clinical  or  technical  personnel 
may  have  set  a  24:00  hour  default.  If  so  the  Duration  screen 
will show the default.  

 
16.   Use  the  UP  &  DOWN  arrow  keys  to  set  the  desired  duration  or  press  YES  to  confirm 

acceptance of the default duration. 

 

Setting the Infusion Rate 
 

17.   The  Pump  calculates  and  displays  the  rate  (in  milliliters  per 

hour) required to deliver the VTBI over the infusion Duration 
confirmed. Rate can be set in increments of 0.1ml between 0.1 
- 100ml/hr and 1ml increments from 100 to 1000ml/hr. 
Senior  clinical  personnel  or  technicians  can  set  a  maximum  rate  limit  to  prevent  rates 
above a certain limit being set, appropriate to the application in use.  
 

18.   Press YES to confirm the calculated rate or use the UP & DOWN arrow to adjust. 

Changing  the  rate  will  alter  the  duration  confirmed  at  the 
previous  step.  In  most  cases  a  maximum  rate  will  be 
programmed into the Pump (see Change Set Up) which will limit 
the  healthcare  provider’s  ability  to  increase  the  infusion  rate 
outside of safe limits. If the rate is changed a Warning screen is 
displayed (right) before the Summary Screen is displayed. 

 

Set Program Lock 
 

19.   Enter the program code and enable the program lock. 

 
Start Infusion 
 

20.  

The Summary Screen confirms Volume To Be Infused, duration 
and  Infusion  rate. 

Healthcare  provider  should    always  check 

that  the  detail  on  this  screen  matches  the    prescription  
especially if they have altered either the default  Duration or 
the calculated Rate during programming. 

Содержание NIKI T34

Страница 1: ...NIKI T34 SYRINGE PUMP Instruction Manual Ref 100 090SS Edition June 2008 0473...

Страница 2: ...Batteries 10 I Access Codes Keypad Lock 11 J Infusion set up programming 12 K Using Accessories with T34 Syringe Pump 16 L Recommended Checks during infusion 17 M Advanced Functions of the T34 18 N Ch...

Страница 3: ...adults or for pediatric use The T34 syringe pump can be easily concealed beneath clothing in a shoulder holster and combination belt A compact lockbox is provided to protect the Pump from damage and o...

Страница 4: ...that unintended programming and key presses do not occur The T34 syringe is contained within a locked box Only the healthcare provider is able to open the box After the healthcare provider programs th...

Страница 5: ...nt may cause serious injury or death Before connecting to the patient purge all air from the syringe and the administration set Depending on the route of administration and the volume of air air embol...

Страница 6: ...e subject it to extreme physical stress This could damage the pump which may impair the operation of the pump and or the accuracy of the infusion If the pump is dropped return the pump for inspection...

Страница 7: ...3 ON OFF key Switches the pump on and off 4 NO STOP key a Takes user back a step b stops pump 5 YES START key Confirms choices 6 Down arrow key scrolls between options 7 Up arrow key Scrolls between...

Страница 8: ...ttery Do not dispose of battery in municipal waste Symbol indicates separate collection for battery is required The use of single use disposable components on more than one patient is a biological haz...

Страница 9: ...Precautions Administration Sets Administration sets must be compatible with the medicine delivered Please check with the manufacturer of the infusion set before use Visually inspect packaging and cont...

Страница 10: ...y and need to be replaced more often Do not use non alkaline or rechargeable batteries There is some variation in size between different brands of batteries Never try to force a battery into the batte...

Страница 11: ...nu used for service and or repair of the Pump is also protected by a special access procedure and different code from Change Set Up Keypad Lock The T34 allows all users to lock the operation of the ke...

Страница 12: ...tus 3 Load the syringe into the Pump prior to connecting the syringe to the patient 4 The LCD display will show PRE LOADING and the actuator will start to move Wait until it stops moving and the syrin...

Страница 13: ...he programmed infusion rate the user can purge the system once only up to a volume limit configured under Change Set Up max 0 5 ml To use this option after syringe confirmation a Press FF key b Ensure...

Страница 14: ...l show the default 16 Use the UP DOWN arrow keys to set the desired duration or press YES to confirm acceptance of the default duration Setting the Infusion Rate 17 The Pump calculates and displays th...

Страница 15: ...START to commence infusion 23 While running the LCD displays infusion Time Remaining top line Infusion Rate main central line and the bottom line will alternate between Syringe Size Brand and Pump De...

Страница 16: ...the T34 place it into the lockbox and carry pouch to protect the syringe Pump syringe contents from damage or tampering 1 Open the lockbox using the standard key that operates all T34 lockboxes 2 Plac...

Страница 17: ...ing the INFO key either a third press or waiting a few seconds returns the display to the main infusion display screen Excessive key presses or usage of the INFO feature will reduce battery life Use o...

Страница 18: ...programming to set a max min rate To titrate the rate up or down during infusion follow these steps Use the UP or DOWN arrows on the keypad to adjust the rate to the desired new setting as displayed...

Страница 19: ...es all events in log are correctly date and time stamped ensure local update for DST FF Key Operation Limits the forward movement of actuator caused by pressing the FF key during syringe loading 0 1 1...

Страница 20: ...Turn pump off Syringe Displaced Check Syringe Press YES to confirm One or more of the syringe detection sensors is not detecting up the syringe Check the syringe and re seat as necessary Pump Paused T...

Страница 21: ...new battery Service required The infusion is going too quickly has ended early or too slowly volume remaining in syringe at end of infusion Incorrect rate set Wrong syringe brand onfirmed during set...

Страница 22: ...ry Pump Paused Too Long Syringe Displaced Syringe Empty T34 Dimensions 169 x 53 x 23mm Classification Type CF Equipment degree of protection against electrical shock Housing ABS fire retardant Weight...

Страница 23: ...he patient and or the patient s assistants or family members in case of questions or concerns Training the Patient and or the Patient s Assistants or Family Members 1 Using the Patient Guide instruct...

Страница 24: ...ith a small dry brush to remove debris or other particles Between Patients Change the battery Before connecting the pump to a patient and periodically during use clean the unit using a lint free cloth...

Страница 25: ...returned to CME or its designated representative should be properly packaged using CME shipping cartons and inserts Inadequate packaging may result in damage to the pump This Limited Warranty shall no...

Страница 26: ...trained certified service centers or biomedical engineers Please consult your local certified engineers or return to your T34 distributor should this need arise For alternate site homecare application...

Отзывы: