10ml BD Plastipak

Volume     12 ml

 Change ↑↓   Press YES 

 Volume       12.0ml
 Duration    24:00
 Rate              0.50ml/h

 Confirm, 

Press YES 

Press YES to Resume
NO for New Program

 

7.  

Finally lower the barrel clamp arm. If the syringe is correctly 
loaded  the  syringe  graphic  (right)  will  become  solid  (no 
flashing  components)  and  the  pump  will  display  the  next 
screen - Size and Brand of the syringe detected. 

 

Syringe Detection & Confirmation 

 
8.  

Check the LCD display to ensure that the Pump has correctly 
identified the syringe size and brand. 
If  the  syringe  brand  displayed  is  not  correct,  use  the  UP  or  DOWN  arrow  keys  to  scroll 
between  brands  of  similar  dimensions.  This  will  only  happen  if  the  syringe  selected  has 
dimensions /-3% of another commonly used brand. 

 
9.  

Press YES/START key to confirm 

 
10.   If the pump was stopped and turned off before the last program 

reached End Program the Resume prompt screen will appear (i.e. 
if,  during  an  infusion,  the  pump  was  turned  off  to  change  the 
battery). Press NO to continue programming the new regime. 

 

11.

 

Purge Option 

 

NOTE

PURGE  OPTION

  –  In  order  to  eliminate/reduce  slack  (visible  spaces  at  the  syringe 

collar & plunger loading points) and ensure a faster start up time (time to reach the programmed 
infusion rate), the user can purge the system (once only) up to a volume limit configured under 
Change Set Up (max 0.5 ml). To use this option after syringe confirmation: 

 

a. Press FF key 
b. Ensure the patient is NOT connected to the set & Press Yes to confirm. 
d. Press and Hold the FF key until the slack is removed and/or purge is completed 
e. Press STOP to return to programming screens when purge is completed. 

 

 

PROGRAMMING MAY NOW TAKE TWO DIFFERENT PATHS DEPENDING ON WHETHER 

THE PUMP HAS BEEN CONFIGURED WITH PROGRAM LOCK ON OR  

OFF.  PROCEED AS FOLLOWS DEPENDING ON THE SCREEN DISPLAYED 

 
 

 
 
 
                                 

PROCEED TO STEP 12

          

          

PROCEED TO STEP 18

 

              

          

PROGRAM LOCK IS OFF                             PROGRAM LOCK IS ON 

 

12.   Once  the  syringe  brand  &  size  are  confirmed  the  pump  calculates  and  displays  the 

deliverable volume in the syringe. 
 

13.   The programmer can use the arrow keys to adjust the Volume To Be Infused (VTBI) e.g. 

to allow a residual volume for KVO if required). The Pump cannot deliver the full contents 
of all syringe brands/sizes so in some cases there may be  a slight  residual volume left in 
the  pump  when  the  actuator  has  traveled  to  the  zero  position.  If  this  is  the  case  the 
deliverable  volume  displayed  will  reflect  this.  For  example,  if  the  visually  confirmed 
contents of the syringe are 18ml the Pump may display 17.5ml (of 18ml) as the deliverable 
volume. In this example 0.5ml will be left in the syringe at End of Infusion. 
 

Содержание NIKI T34

Страница 1: ...NIKI T34 SYRINGE PUMP Instruction Manual Ref 100 090SS Edition June 2008 0473...

Страница 2: ...Batteries 10 I Access Codes Keypad Lock 11 J Infusion set up programming 12 K Using Accessories with T34 Syringe Pump 16 L Recommended Checks during infusion 17 M Advanced Functions of the T34 18 N Ch...

Страница 3: ...adults or for pediatric use The T34 syringe pump can be easily concealed beneath clothing in a shoulder holster and combination belt A compact lockbox is provided to protect the Pump from damage and o...

Страница 4: ...that unintended programming and key presses do not occur The T34 syringe is contained within a locked box Only the healthcare provider is able to open the box After the healthcare provider programs th...

Страница 5: ...nt may cause serious injury or death Before connecting to the patient purge all air from the syringe and the administration set Depending on the route of administration and the volume of air air embol...

Страница 6: ...e subject it to extreme physical stress This could damage the pump which may impair the operation of the pump and or the accuracy of the infusion If the pump is dropped return the pump for inspection...

Страница 7: ...3 ON OFF key Switches the pump on and off 4 NO STOP key a Takes user back a step b stops pump 5 YES START key Confirms choices 6 Down arrow key scrolls between options 7 Up arrow key Scrolls between...

Страница 8: ...ttery Do not dispose of battery in municipal waste Symbol indicates separate collection for battery is required The use of single use disposable components on more than one patient is a biological haz...

Страница 9: ...Precautions Administration Sets Administration sets must be compatible with the medicine delivered Please check with the manufacturer of the infusion set before use Visually inspect packaging and cont...

Страница 10: ...y and need to be replaced more often Do not use non alkaline or rechargeable batteries There is some variation in size between different brands of batteries Never try to force a battery into the batte...

Страница 11: ...nu used for service and or repair of the Pump is also protected by a special access procedure and different code from Change Set Up Keypad Lock The T34 allows all users to lock the operation of the ke...

Страница 12: ...tus 3 Load the syringe into the Pump prior to connecting the syringe to the patient 4 The LCD display will show PRE LOADING and the actuator will start to move Wait until it stops moving and the syrin...

Страница 13: ...he programmed infusion rate the user can purge the system once only up to a volume limit configured under Change Set Up max 0 5 ml To use this option after syringe confirmation a Press FF key b Ensure...

Страница 14: ...l show the default 16 Use the UP DOWN arrow keys to set the desired duration or press YES to confirm acceptance of the default duration Setting the Infusion Rate 17 The Pump calculates and displays th...

Страница 15: ...START to commence infusion 23 While running the LCD displays infusion Time Remaining top line Infusion Rate main central line and the bottom line will alternate between Syringe Size Brand and Pump De...

Страница 16: ...the T34 place it into the lockbox and carry pouch to protect the syringe Pump syringe contents from damage or tampering 1 Open the lockbox using the standard key that operates all T34 lockboxes 2 Plac...

Страница 17: ...ing the INFO key either a third press or waiting a few seconds returns the display to the main infusion display screen Excessive key presses or usage of the INFO feature will reduce battery life Use o...

Страница 18: ...programming to set a max min rate To titrate the rate up or down during infusion follow these steps Use the UP or DOWN arrows on the keypad to adjust the rate to the desired new setting as displayed...

Страница 19: ...es all events in log are correctly date and time stamped ensure local update for DST FF Key Operation Limits the forward movement of actuator caused by pressing the FF key during syringe loading 0 1 1...

Страница 20: ...Turn pump off Syringe Displaced Check Syringe Press YES to confirm One or more of the syringe detection sensors is not detecting up the syringe Check the syringe and re seat as necessary Pump Paused T...

Страница 21: ...new battery Service required The infusion is going too quickly has ended early or too slowly volume remaining in syringe at end of infusion Incorrect rate set Wrong syringe brand onfirmed during set...

Страница 22: ...ry Pump Paused Too Long Syringe Displaced Syringe Empty T34 Dimensions 169 x 53 x 23mm Classification Type CF Equipment degree of protection against electrical shock Housing ABS fire retardant Weight...

Страница 23: ...he patient and or the patient s assistants or family members in case of questions or concerns Training the Patient and or the Patient s Assistants or Family Members 1 Using the Patient Guide instruct...

Страница 24: ...ith a small dry brush to remove debris or other particles Between Patients Change the battery Before connecting the pump to a patient and periodically during use clean the unit using a lint free cloth...

Страница 25: ...returned to CME or its designated representative should be properly packaged using CME shipping cartons and inserts Inadequate packaging may result in damage to the pump This Limited Warranty shall no...

Страница 26: ...trained certified service centers or biomedical engineers Please consult your local certified engineers or return to your T34 distributor should this need arise For alternate site homecare application...

Отзывы: