I – Access Codes & Keypad Lock
Secure Provider Access Codes
The T34 has three levels of healthcare provider access protection to prevent unauthorized
changes to the set up, configuration or program. Although no access code is required to turn the
pump on and run an infusion, certain settings and features may be configured and locked based
on patient need.
No access codes are contained in this manual. Please refer to senior clinical or technical staff for
these codes if you have been trained and certified in their use. Do not share access codes with
unauthorized personnel.
Program Lock
Always use the Program Lock when the pump is used in homecare environments to prevent the
patient changing the prescription.
Technicians Menu
‘Technicians Menu’, used for service and/or repair of the Pump, is also protected by a special
access procedure and different code from ‘Change Set Up’.
Keypad Lock
The T34 allows all users to lock the operation of the keypad during infusion to prevent
tampering with the keypad if program lock is not used and prevent inadvertent power off if
program lock is used. To active the keypad lock: -
with the pump infusing, press and hold the INFO key until a chart is displayed showing a bar
moving from left to right.
Hold the key until the bar has moved completely across the screen and a beep is heard to
confirm the lock has been activated.
To turn the lock off repeat this procedure. The bar will now move from right (ON) to left
(OFF) and a beep will be heard to confirm.
The keypad lock does not affect the operation of the STOP keys as patients unaware of how to
deactivate the lock may need to STOP an infusion in an emergency (for example, when clearing
an alarm state).