2008T Hemodialysis Machine with CLiC Device User’s Guide
P/N 490206 Rev E
33
Crit-Line in a Clip Specifications
Dimensions
1.75”H, 1.00”W, 3.00”L
Weight 0.25
lbs.
Cable length
42”
Operating Temperature
50 °F to 104 °F (10 °C to 40 °C)
Unit will be warm to the touch during use
(max. 55.3 °C @ 40 °C ambient)
Storage and Transportation Temperature
-40 °F to 257 °F (-40 °C to 125 °C)
Oxygen Saturation Instrument Range &
Accuracy
@ Hct
Accurate
within ± 3%
Accurate
within ± 5%
45 – 60
60 – 100
50 – 100
20 – 45
50 – 100
30 – 100
10 – 20
Not Specified 40 – 100
Hematocrit Instrument Range & Accuracy
10 Hct – 60 Hct: ± 1 Hct SD
Power
5 VDC / MAX 0.36A, USB bus powered
Water Ingress Protection
IPX0 (CLiC device)
IPX1 (2008T hemodialysis machine)
Anesthetic Suitability Not
suitable
Modes of Operation
Continuous or Standby
Note
: The CLiC device is a Clinical Laboratory Improvement Amendment (CLIA) exempt
instrumentation device