2008T Hemodialysis Machine with CLiC Device User’s Guide
P/N 490206 Rev E
13
Reading Graphs on the “Crit-Line” Screen
The “Crit-Line” screen can display either Blood Volume (BV) or
Hematocrit (Hct) on the upper graph depending on the Service Mode
setting.
On the lower graph, the Oxygen Saturation (O
2
Sat) graph can be changed
to display blood pressures by selecting the graph and pressing the
CONFIRM
key.
The BV and O
2
Sat graphs are displayed by default.
Graphing of the data begins after the CLiC device reads blood sensed and
the hematocrit has been stable for sixty seconds with the Tx Clock and
blood pump running. Graphs are displayed for a minimum of four (default)
and a maximum of ten hours. The Hct and O
2
Sat graphs are resized
vertically during the treatment depending on the min/max values and alert
levels.
Listed below are features common to all graph displays:
Feature
Function
Elapsed Time – The time the Tx Clock and CLiC device are
running. If the Tx Clock is paused or an event pauses the
treatment (such as blood alarms), then the Elapsed Time is
paused along with CLiC device data collection.
Initial Hct – The initial hematocrit (Hct) value is recorded
along with the first data point when graphing begins.
Initial est. Hb – The estimated initial hemoglobin (Hb) value
is recorded along with the first data point when graphing
begins. It is calculated based on the hematocrit reading.
Markers
button – Selecting this button displays a menu to
insert an event marker on the BV (or Hct) and O
2
Sat graphs.
Use the
or
(up or down) keys on the keyboard to select
either ‘Symptom’ or ‘Intervention’. Pressing the
CONFIRM
key places the marker and the current
∆
BV%, Hct, and
oxygen saturation values on the latest point on the graph;
pressing the
Escape
key exits the menu without placing a
marker.
The
Symptom
marker is displayed as a yellow diamond
The
Intervention
marker is displayed as a black
triangle
Note
: Markers can only be set at a minimum of ten minute
intervals (depending on graph size) and when the Tx Clock
is running and the CLiC device is not disabled. Resizing the
graph may cause markers to rotate in order to better fit the
space depending on the limits. If there is not enough space
between markers, the machine will not place a marker in
order to prevent overlap.