2008T Hemodialysis Machine with CLiC Device User’s Guide
P/N 490206 Rev E
11
Monitoring the hemodialysis treatment from the “Crit-Line” screen
When the CLiC device is used for treatment with the 2008T hemodialysis
machine, the “Crit-Line” screen replaces the “BTM BVM” screen. Most of
the commonly viewed data from other screens are grouped on the “Crit-
Line” screen for convenient monitoring.
Listed below are the features of the “Crit-Line” screen that can also be
found on other screens.
Feature
Description
Art Press
Arterial Pressure – The value displayed on the “Home” screen
(mmHg)
Ven Press Venous Pressure – The value displayed on the “Home” screen
(mmHg)
TMP
TMP (Transmembrane Pressure) – The value displayed on the
“Home” screen (mmHg)
UF Goal
UF Goal – The UF goal selected for the treatment and set on the
“Home” screen (ml)
UF Rate
UF Rate – The UF rate selected for the treatment and set on the
“Home” screen (ml/h)
UF Rmvd
UF Removed – The UF removed at a given point in the treatment
and displayed on the “Home” screen (ml)
Dial Flow
Dialysate Flow – The dialysate flow rate selected for the treatment
and set on the “Home” screen (ml/min). The dialysate flow rate can
also be turned on or off from the “Dialysate” screen.
Temp
Temperature – The temperature of the dialysate as displayed on
the “Home” screen (°C)
Cond
Conductivity – The conductivity of the dialysate as displayed on the
“Home” screen (mS/cm)
Blood Flow Blood Flow Rate – The current blood pump rate (ml/min)
Hep Rate
Heparin Rate – The heparin pump rate set on the “Heparin” screen
(ml/h)
Note
: As on the “Heparin” screen, this value will flash when the
heparin pump is off unless both the “Heparin” screen
Rate
and
Infusion Time
buttons are set to 0 (zero).
Kecn
Kecn – The current Kecn (amount of sodium clearance)
measurement. The Kecn is also displayed on the “Kt/V AF”
screen’s “OLC Data” subscreen.
Treatment (Tx) Clock – This button runs or pauses the treatment
and displays the Remaining Time of Dialysis (RTD). The green
segment of the pie chart represents the amount of treatment
completed and grows as the treatment progresses. A
Tx Clock
button and an
RTD
button to set the RTD are also located on the
“Home” screen.
For a full description of these buttons, see the
2008T Hemodialysis
Machine Operator’s Manual
(P/N 490122).