2008T Hemodialysis Machine with CLiC Device User’s Guide
P/N 490206 Rev E
9
Connecting the CLiC Device to the Blood Chamber
Follow the steps below to clip the CLiC device to the Crit-Line Blood
Chamber (P/N CL10041021) before connecting to the patient’s vascular
access:
1 Pinch open the CLiC device
Hold the CLiC device such that one side is held by the index finger and
other side held by the thumb. Squeeze the clip to spread the sensor
elements apart.
2 Clip CLiC device on Blood Chamber
Caution
: Before attaching the CLiC device, check carefully for
leaks at the connections between the blood chamber and the
dialyzer and the blood chamber and the arterial blood line.
Improper attachment of the arterial bloodline to the Crit-Line
Blood Chamber can cause blood or saline to leak onto and into
the CLiC device. This can damage the CLiC device.
Note
: Make sure that the CLiC device is plugged into the USB
port located on the right side of the 2008T hemodialysis machine
display screen.
1.
Place the CLiC device over the Crit-Line Blood Chamber so that the
optical sensor covers the lens of the Crit-Line Blood Chamber.
2.
Release the clip to allow the optical sensor to seat on the viewing area
of the Crit-Line Blood Chamber.
3.
Make certain that the CLiC device has locked securely into place on the
Crit-Line Blood Chamber by noting the audible and tactile click.
4.
Ensure that the CLiC device is perpendicular to the top of the dialyzer
and cannot be easily rotated up or down while attached to the blood
chamber to ensure its proper placement.
Note
: Make sure that the CLiC device is properly in place before
starting the treatment. If the CLiC device is not properly seated
on the Crit-Line Blood Chamber, the accuracy of the hematocrit
and oxygen saturation readings may be adversely affected.
3 Check for proper blood flow before starting the Tx Clock
When starting the blood pump at the beginning of treatment, visually
verify proper blood flow in the extracorporeal circuit, including the Crit-
Line Blood Chamber, before starting the Tx Clock.