27
P/N 49020
6 Rev E
2008T
Hem
odialysi
s Mach
ine with CLiC Device
User’
s Guid
e
Message
Purpose o
f Message
Ty
pe
Action Required
Crit-Line: No
Blood
The Tx Clock is running and
the CLiC de
vice no longer
senses blood in the Crit-Line
Blood Cham
ber.
Warnin
g
Check the Crit-Line Blood Chambe
r:
Make sure there is proper
blood flow: There should
be no air or
bubbles in the chamber.
Make sure the CLiC de
vice is
properly
seated on the Crit-Line Blood
Chambe
r: The device sh
ould be plac
ed per
pendic
ularly o
ver th
e
chamber and
cannot be ro
tated up or d
own.
If
the messa
ge is not cle
ared, the
CLi
C
device can
not be used for the
dialysis treatment.
Remove the device from the Crit-Lin
e Blood Cham
ber.
Disconnect the CLiC de
vi
ce cable
from the USB p
ort and press the
RESET
key to disable the CLiC device
.
Contact a qualified
service technician.
Crit-Line: No
Comm
The 2008T h
emodialysis
machine has not received
data from the CLiC de
vice.
Warnin
g
Check the CLiC device’s USB cable:
Make sure the cable is
securely conne
cted to the 2
008T
hemodial
ysis machine’s U
SB port.
Wait up to one minute for the messag
e to clear.
If
the messa
ge is not cle
ared, the
CLi
C
device can
not be used for the
dialysis treatment.
Disconnect the CLiC de
vi
ce cable from the USB p
ort.
Press the ‘Y’ key on the Keybo
ard
to di
sable the CLiC device.
Contact a qualified
service technician.
Crit-Line: Ob
struction
Something is blocking the
CLiC de
vice’s optical
sensor.
Warnin
g
Check the Crit-Line Blood Chambe
r:
Remove the CLiC de
vice from the blood chamber a
nd check for
obstructions.
Reattach the CLiC device
to the blood chamber.
Wait up to one minute for the messag
e to clear.
If
the messa
ge is not cle
ared, the
CLi
C
device can
not be used for the
dialysis treatment.
Remove the CLiC de
vice from the Crit-Line Blood C
hamber.
Disconnect the CLiC de
vi
ce cable from the USB p
ort.
Press
the
R
ESET
key to
disable the CLiC device.
Contact a qualified
service technician.
Crit-Line o
n Filter?
To verify the CLiC de
vice, the
device must
be clipped to its
ve
rification fil
ter.
Advisory
message
Place the CLiC device on
the CLiC Ver
ification Filter, which is a
ttached to
its USB cable.