2008T Hemodialysis Machine with CLiC Device User’s Guide
P/N 490206 Rev E
7
Connecting the Crit-Line Blood Chamber
Follow these basic steps to attach the disposable Crit-Line Blood Chamber
(P/N CL10041021) to the dialyzer during the extracorporeal bloodline
setup before priming the bloodlines.
Note
: The Crit-Line Blood Chamber is intended for single use
only. A new sterile blood chamber must be used for each
monitoring session.
1 Inspect Crit-Line Blood Chamber and packaging
Inspect the Crit-Line Blood Chamber and its sterile package prior to use.
Refer to the blood chamber package label to ensure that the blood chamber
sterilization has not expired.
Warning
: Do not use the Crit-Line Blood Chamber if its
package has been opened or its sterility has otherwise been
compromised prior to use. The viewing area of the disposable
blood chamber should be kept clean and free of obstruction.
2 Connect Crit-Line Blood Chamber to dialyzer
Remove the blood chamber from its sterile package and aseptically attach
it to the arterial port of the dialyzer (as shown on the facing page). Make
sure the connection is tight.
3 Continue setting up the bloodlines
Continue to set up the bloodlines as described in Chapter 2, “Connecting
the Extracorporeal Blood Circuit,” in the
2008T Hemodialysis Machine
Operator’s Manual
(P/N 490122).
When connecting the arterial bloodline to the Crit-Line Blood Chamber
(not the dialyzer), hold the Crit-Line Blood Chamber securely in one hand
and aseptically attach the dialyzer end of the arterial bloodline with the
other hand. Be careful that you do not cross-thread the connection. The
connection of the Crit-Line Blood Chamber to the arterial bloodline must
not leak.
Note
: Make sure that no air is in the Crit-Line Blood Chamber
after priming. Any air present in the Crit-Line Blood Chamber
will cause the hematocrit reading to be inaccurate.