27
Immunity Guidance Table
Guidance and manufacturer’s declaration
–
Electromagnetic Immunity
The HiFlo2 suction pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the HiFlo2 suction pump should ensure that it is used in such an
environment.
Immunity
Test
Level indicated by
EN 60601
-
1
-
2
Compliance
Level
Electromagnetic environment
-
guidance
Electrostatic
discharge
(ESD)
EN 61000
-
4
-
2
± 8 kV on contact
± 15 kV in air
The device
doesn’t
change its
state
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient / burst
EN 61000
-
4
-
4
± 2 kV power supply
lines
± 1 kV for input /
output lines
The device
doesn’t
change its
state
Mains power quality should be that
of a typical commercial
environment or hospital.
Surge
EN 61000
-
4
-
5
± 1 kV differential
mode
The device
doesn’t
change its
state
Mains power quality should be that
of a typical commercial
environment or hospital.
Loss of voltage,
brief voltage
interruptions
and variations
EN 61000
-
4
-
11
5%U
T
(>95% dip U
T
)
for 0.5 cycle
40%U
T
(>60% dip U
T
)
for 5 cycle
70%U
T
(>30% dip U
T
)
for 25 cycle
<5%U
T
(>95% dip U
T
)
for 5 sec
-
Mains power quality should be that of
a typical commercial environment or
hospital. If the user of the HiFlo2
suction pump requires that the
appliance operates continuously, the
use of a backup supply is
recommended.
Magnetic field
EN 61000
-
4
-
8
30 A/m
The device
doesn’t
change its
state
The power frequency magnetic field
should be measured in the
intended installation location to
assure that it’s sufficiently low.
Note U
T
is the nominal value of the power supply voltage
Содержание SUC84602