Rinse with hot running water and dry all parts with a soft non-
abrasive cloth. It is possible to wash with commercial disinfectants by
carefully following the instructions and dilution values supplied by
the manufacturer. Do not use phenolic solutions as disinfecting
agents in polycarbonate jars.
After cleaning, leave the parts to dry in an open, clean environment.
Suction Tubing
The silicone aspiration tubes and the conical fitting may be carefully
washed in hot water (temperature must not exceed 60°C). After clean-
ing, leave the parts to dry in an open, clean environment.
Suction Tubing may be sterilised with sterilants determined by local
protocols and compatible with silicone tubing. Suction tubing may be
autoclaved to a maximum of 121°C for 15 minutes. At higher temper-
atures the suction tubing will discolour and lose shape.
Autoclaving
Separate the jar and lid items as detailed in the cleaning section
above.
The following items may be autoclaved in a single steam sterilisation
cycle at 121°C and 1 bar for 15 minutes:
•
Jar - position jar upside down in autoclave
•
Lid - position lid upside down in autoclave
•
Silicone tubing
•
Conical connector
•
Trap jar - position jar upside down in autoclave
Re-Assembly
Place float valve into overflow cage with the seal facing up. Connect
overflow cage to lid and lock bayonet on float cage by rotating cage
anticlockwise
. Place lid seal carefully into groove in lid. Fit lid to jar
and lock securely into place by engaging bayonet and rotating lid
firmly clockwise.