Suction Accessories
6.
Occlude tubing and adjust vacuum control to the required
vacuum level. (Clockwise = increase)
Always reduce the vacuum level for paediatric and neonatal
patients.
7.
Apply suction to patient with a compatible suction cannula
fitted to the patient tubing.
8.
When finished, switch off pump by pressing switch.
Ensure that the mains power plug is accessible at all times when
pump is in use, in case a positive disconnection from the mains
power supply is required.
Suction Accessories
The device and its accessories are biocompatible in accordance with
EN 60601-1.
Suction Cannulae
Suction cannulae or suction probes for contact with the human body
should comply with ISO 10993-1 requirements for biocompatibility.
Collection Jar (Canister)
The mechanical strength of the reusable collection jar (canister) is
guaranteed for 30 cycles of cleaning and sterilization. Beyond this,
there may be signs of decay and replacement is recommended.
Silicone Tubing
The number of cleaning and sterilization cycles of the silicone tubing
is dependent on the actual usage. Tubing should be checked for
cracking and other visible signs of wear before re-use.