Waste Disposal
Check Bacteria Filter
The bacterial/hydrophobic filter is an important device to help protect the pump
and environment from moisture and bacteria from the patient. It is
recommended that the filter be changed every month or when damp or
discoloured. Do not clean, sterilise or disassemble filter.
Change filter after every use when used on infectious patients
or patients whose infection status is unknown.
After Every 100 Hours or 2 Months of Operation
•
Check all suction tubing and replace if it is perished, soft or
discoloured.
•
Check the seal ring on the lid and replace if hard, cracked or
perished. Check the fit of the lid in the collection jar (canister).
•
Check the overflow cut off valve seal and replace if perished or
damaged.
Waste Materials
The contents of the collection jars, trap jars, suction tubing and
bacteria filter may contain biohazard wastes. Handle using safe
handling procedures, which may include the use of rubber gloves and
eye protection, and dispose of according to local protocols for
biohazard materials.
Recycling
At the end of their service life, the pump and accessories should be
dismantled if necessary, and disposed of according to the WEEE
directive.