15
Using FLOVAC Liners
1. Fit FLOVAC ring bracket to multipurpose rail.
2. Insert liner bag [1] into the matching size
support jar [2].
Press edge of lid firmly all around edge to
ensure seal.
3. Close the TANDEM inlet [3] with the attached
cap [4].
4. Fit jar assembly into FLOVAC ring bracket.
5. Press short side of yellow connector firmly into
matching yellow port [5] on the lid.
6. Using a short length of tubing connect the
yellow connector to centre port of the trap jar.
7.
Connect patient tubing to white inlet elbow[7].
Connect patient tubing to suitable suction
cannula.
8. Follow steps in Operation section.
Using FLOVAC Disposable Liners
9
1
2
7
4
3
5
If fluid reaches the maximum level the hydrophobic filter in the
lid will operate to prevent further fluid flow. When this
happens switch off pump and replace liner.
Disposal
1.
Switch off pump.
2.
Disconnect tubing from jar.
3.
Remove the white elbow [7] and close the inlet port with
the cap [9] attached to lid.
4.
Using the lid handle remove the liner [1] from the support
jar and dispose of liner according to hospital protocol.
5.
Retain support jar [2] for reuse.
Support Jar Cleaning
The reusable support jar may be cleaned either with water and
neutral detergents or autoclaved at 121
°
C for 15 minutes.
Do not use solvents or alcohol for cleaning and disinfection.
These may damage the support jar.
The mechanical integrity of the support jar is guaranteed for
30 cleaning cycles under the specified conditions.
After this, the jar may show signs of wear and replacement is
recommended.