Specification
43
Flash Test (CLASS 1)
Test
Voltage
1.25kV
a.c
Display Range
0.1 - 10 mA a.c.
Resolution
0.01mA
Accuracy
+/- 5% +/- 2 digits
Pass Levels
selectable up to 10mA
Flash Test (CLASS2)
Test
Voltage
2.5kV
a.c
Display Range
0.1 - 10 mA a.c.
Resolution
0.01mA
Accuracy
+/- 5% +/- 2 digits
Pass Levels
selectable up to 10mA
Leakage
Test
Voltage
Mains
Supply
Display Range
0.1 - 20.0 mA a.c.
Resolution
0.01mA
Accuracy
+/- 10% +/- 2 digits (1.00 - 20mA)
Pass Levels
selectable up to 10mA
Power Tests
Resolution
0.01kVA
Measured Load
0 - 4kVA
Accuracy
+/- 10 of reading, +/- 2 digits
IEC Lead Test
Test
Voltage
40Vac
Detects
Good,
Open,
Short,
Reversed
Mechanical
Size
450mm x 410mm x 155mm
Weight
3.5kg