Selecting The Test Settings
13
Earth Bond Test
Current = 25A
Duration = 2 seconds
Pass level = below 0.5
Ω
Line Short to Case
Voltage = 40Vac
Duration = 2 seconds
Pass level = below 1mA
Flash Test
Voltage = 1250Vac
Duration = 2 seconds
Pass level = below 3mA
Power/Leakage Test
Voltage = Supply voltage
Duration = 5 seconds
Power Pass level = Nominal >=4
Ω
Leakage Pass level = below 1mA
B:
DOUBLE INSULATED / CLASS 2
This test selection is for domestic and hand tool products that have
a plastic or insulated case and that require a full range of tests
including Flash. The standard test settings for this selection are:
Line Short to Case
Voltage = 40Vac
Duration = 2 seconds
Pass level = below 1mA