Selecting The Test Settings
14
Flash Test
Voltage = 2500Vac
Duration = 2 seconds
Pass level = below 2mA
Power/Leakage Test
Voltage = Supply voltage
Duration = 5 seconds
Power Pass level = Nominal >=4
Ω
Leakage Pass level = below 1mA
C: POWER CORD TEST
This test selection is for the testing of detachable power cords such
that used by computers & kettles etc. The standard test settings for
this selection are:
Earth Bond Test
Current = 100mA
Duration = 2 seconds
Pass level = below 0.5
Ω
Line Short to Case
Voltage = 40Vac
Duration = 2 seconds
Pass level = below 1mA
Flash Test
Voltage = 1250Vac
Duration = 2 seconds
Pass level = below 3mA