Specification
42
Specification
Earth Bond Test
Test Voltage
6V nominal (no load)
Test Current
4A, 10A, 25A (into s/c load)
Range
40m
Ω
- 19.99
Ω
Resolution
1m
Ω
Accuracy
+/- 5% +/- 2 digits
Pass
Levels
user
selectable
Earth Screen Test (IT Earth Bond Test)
Test Voltage
100mV nominal (no load)
Test Current
100mA (into s/c load)
Range
40mW
-
19.99W
Resolution
1m
Ω
Accuracy
+/- 5% +/- 2 digits
Pass
Levels
user
selectable
Insulation Test
Test
Voltage
500V
Display Range
0.1 - 99.99M
Ω
Accuracy
+/- 5% +/- 2 digits (up to 19.99M
Ω
)
Short to Case Test
Test Voltage
40V a.c. o/c
Display Range
0.1 - 20.0 mA a.c.
Resolution
0.01mA
Accuracy
+/- 10% +/- 2 digits (1.00 - 20mA)
Pass Levels
selectable up to 15mA