How to use Setup
29
POWER POLARITY
This option will allow the tester to perform the power tests in one of
three ways.
INPUT POLARITY - perform the power tests with the same mains
polarity as the Safe Check 8 is powered.
AUTO CORRECTED - perform the power tests with the mains
correct polarity even if the Safe Check 8 is powered from a socket,
which has the mains polarity, reversed.
INPUT & REVERSED - this option will perform two power tests for
every one selected, one with the correct mains polarity and one with
the reverse mains polarity.
Default is INPUT POLARITY.
ON TEST FAILURE
This gives the User the option of having a menu for a failure within
an automatic sequence. This menu will give you the option of
retrying a failed test or appliance.
Default is FAIL MENU.
AFTER TEST
The options set here determine the next action after all the test
results are saved to memory. Options are: -
OPTIONS – Displays the OPTIONS menu giving the User further
choices.
NEW TEST – Will display the ENTER APPLIANCE DETAILS screen
to immediately commence the next test.
DOWNLOAD, NEW TEST – Will Download the test results with out
further prompting and display the ENTER APPLIANCE DETAILS
screen to immediately commence the next test. When this option is
chosen a further option is required to CONFIGURE the Download.
Default is OPTIONS.