Hydrogen generator
Hydro200
9
4.
Installation
4.1.
Installation area requirements
4.1.1.
Humidity and dust
To avoid risks of damage to electronic components, install the generator in an
environment subject to limited relative humidity and low concentrations of dust.
The generator must also be protected against drips, rain and wind.
4.1.2.
Temperature
The ambient temperature in the generator installation area must be between
5°C and 40°C.
Install away from heat sources. Avoid direct exposure to sunlight.
4.2.
Installation procedure
4.2.1.
Step 1: remove the packaging
Remove all packaging, taking care not to damage the generator panels.
If possible, store the packaging to ensure adequate protection when moved in
the future.
Ensure that the supply is complete (see section 3.1)
4.2.2.
Step 2: position the generator
Locate an installation area that meets the specifications listed in section 4.1.
Position the generator on a laboratory bench or on a Claind generator from the
series AirToc, AirToc SL, or NitroAir.
e
Handle with care: never tilt the generator and avoid impact with obstacles.
e
Ensure a clearance of at least 20cm cm at the rear of the generator.
e
Remember never to place objects on the top panel of the generator.
4.2.3.
Step 3: fill the demineralised water tank
• locate the demineralised water tank at the rear of the generator
• disconnect the tank by means of the quick coupling at the bottom of the tank.
• remove the tank
• unscrew the cap, fill the tank to 1 cm from the top edge and refit the cap
• refit the tank and connect the quick coupling
e
Use demineralised water with electrical conductivity no higher than