Use
16
6.5.
Generator shutdown
6.5.1.
Correct shutdown
Correct shutdown of the generator starts from the condition of
GENERATOR
READY
. par. 6.4. describes how to reach this condition.
Set the "POWER" key (black) to OFF to turn off the generator.
6.5.2.
Incorrect shutdown
Avoid turning off the generator directly, without first shutting down as described
in par. 6.4.
However, this may occur inadvertently or due to a power failure. In this case, on
subsequent generator restart, the following message is displayed:
WARNING
INCORRECT SWITCH OFF
To shut off the alarm, press the function key
.
6.6.
Parameter display
See also flow chart in APPENDIX
6.6.1.
Hydrogen and oxygen pressure display
During production, the main menu displays the message
PRODUCTION
.
For more information on generator status, press the up arrow key
. The fol-
lowing page is displayed:
H2
O2
X.XX bar
Y.YY bar
where:
X.XX
= hydrogen compartment pressure bar
Y.YY
= oxygen compartment pressure bar
The values are updated continuously to verify correct operation
of the generator
To exit this page and return to the main menu press up arrow key
.
6.6.2.
Maintenance counters display
From the main menu (
PRODUCTION
), press the down arrow key
:
The following information is displayed in sequence (approx. every 5 seconds)