Hydrogen generator
Hydro200
11
e
If automatic start-up of the generator is necessary, this command should be en-
tered in the set-up program
4.2.6.
Step 6: Connect the vent hose
• locate the outlet O2 VENT at the rear of the generator
• remove the plastic cap using a 12mm hex key
The O2 VENT outlet enables the oxygen (produced in parallel to the generation
of hydrogen) to be released into the atmosphere. As the released oxygen may
in some phases contain traces of electrolyte, the vent should be conveyed as fol-
lows:
• connect the polyamide pipe 6x1 mm (supplied with the generator) and route the
other end of the pipe (shortened if necessary) to a small collection tray placed
at a lower level than that of the generator.
e
avoid obstructing the vent: during use, the end of the polyamide pipe 6x1 mm
opposite the generator must never by closed off or immersed in the liquid.
e
conserve the cap, as it is used in the event of future transport of the generator
(see chapter 5)
4.2.7.
Step 7: set the mode parameters
In this phase the mode parameters should be checked (and modified if requi-
red), which enable the selection of the language for message display and selec-
tion of a number of operating modes, including the method for demineralised
water refilling.
For details, refer to section 6.7.1.
4.3.
Packaging disposal
All packaging materials used are compatible with the environment and may be
stored without risk, or disposed of and recycled in accordance with local legisla-
tion governing classified waste collection.
However, when possible, it is recommended to store the packaging to ensure
adequate protection when moved in the future.