Data Output Examples
+ 123.4567 g
Position
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
+
1
2
3
.
4
5
6
7
g
LF CR
or
+
6
1
7
.
2
8
3
5
c
t
LF CR
or
N
1
+
2
0
.
0
0
0
0
g
LF CR
or
T
o
t
+
6
0
.
0
0
0
0
g
LF CR
Position 1 - 4
:
ID code Character or Space
Position 5
:
Sign
Position 6 - 20
:
Weight with Decimal Point; leading zeros = space
Position 21
:
Space
Position 22 - 24
:
Unit Symbol or Space
Position 25
:
Line Feed
Position 26
:
Carriage Return
109
110
You can connect a computer to your balance to send commands via the balance interface port to control balance functions
and applications.
Format for commands
[
Command Code
Data
]
•[: it shows start of command frame.
•Command Code: it shows which functionality to be carried out for this command frame.
•Data: This field in frame is optional and it is intended to provide data information between Bi-directional communications.
•]: it shows end of command frame.
Commands
[W] : If host computer send this command to balance then balance will Send weight with current unit.
[T] : If host computer send this command then balance will do taring in balance. If stability is not achieved within 45 second
then it comes to that specific feature till that time it shows “-------“on LCM.
9.2 Data Input Format
ID code
ID Code
characters
Meaning
Characters
Meaning
nRef
Counting: Reference sample quantity
N1
Formulation,
Net : N1
wRef
Piece Counting, Percentage Weighing :
N
Formulation,
Net N
Qnt
Piece Counting: Quantity
Tot
Formulation,
pRef
Weighing in percent: Reference percentage
Pur
density : Purity of Gold
Pct
Weighing in percent: Reference percentage
Den
Density : density of sample
Cnt
Animal weighing: No. of sub-weighing operations
Pip
Pipette Calibration
xNt
Animal weighing: Calculated average
Sta
Statistics : to obtain the statistics of the data
Totalization
Totalization
Reference weight
Totalization : Total Weight