37
5.2.3 Adjusting the stability Filter
You can use the stability Filter to match the balance to the ambient conditions.
2 Setting with normal balance surroundings (factory setting)
3 Setting with unstable balance surroundings. The balance operates slower
but is less sensitive to external influences (vibrations, etc.)
4 Setting with extreme unstable balance surroundings. The balance operates even
slower but is less sensitive to external influences (vibrations, etc.)
1 Setting with very stable balance surroundings. The balance operates very quickly
but is sensitive to external influences (vibrations, etc.)
5.2.4 Selecting Calibration Option
User can select any of the Two option for Calibration.
CAL ET
If the user select this option then the machine will perform External Calibration
when the CAL key is press & hold to display “CAL Et” & at this moment if user
release the key, user can enter into the External calibration.
CAL OFF When user press & hold CAL key, directly “Menu” appears on the display
without CAL Et option. Thus user cannot enter into the calibration process.
CAL INT If the user select this option then the machine will perform External Calibration
when the CAL key is press & hold to display “CAL Int” & at this moment if user
release the key, user can enter into the Internal calibration.
5.2.5 Calibration Test
User Can select any Two calibration test option.
CAL T ON If the user select this option then the machine will perform Calibration Test
when the CAL key is press & hold Calibration Test will perform on Internal
or External which ever is selected in calibration menu.
CAL T OFF Actual Calibration will be performed When the CAL key is press & hold .
5.2.6 Auto Zero Tracking
In this option, user can select
whether to enable or disable Auto Zero Tracking
(Factory setting is ON)
The auto zero tracking continuously corrects any deviation from the zero point for
example which can be caused due to slight contamination (i.e. due to dust particles)
on the weighing pan.
38