89
With this feature, the user can obtain the statistics of the data stored in the scale. These
statistics includes the details such as number of readings 'n', Average, Minimum value,
Maximum value, Standard deviation, Difference and Co efficient of Variance.
The feature 'F StAt' can be activated by selecting the 'F StAt' function in the menu. This can
be done by pressing the <CAL/MENU> key for 4 seconds and changing the feature by
<TOGGLE> key. Press <TARE> key to select 'F StAt'. press <TARE> for 2 seconds to store the
selection.
The below procedure presupposes that you have selected the 'F StAt' function in the user
menu.
Procedure
ð
Place the empty container on the pan.
ð
Press the < TARE> key briefly to tare the balance.
ð
Press the <TOGGLE> key for entering the 'F StAt' function.
Note:
To use the 'F StAt' function, it is assumed that the user has already performed the above
procedure and the scale is in 'F StAt' mode.
Statistics procedure
ð
The user can now place weight on the pan and press the <CAL/MENU> key once
the stability is achieved.
ð
The scale should display 'n-1' (where 'n' is the number of the current weight) and will
retain this as first weight.
ð
“rmv” will be displayed on screen for 2 seconds to instruct the user to unload the
weight.
ð
The next weight will be taken into statistics only after user has taken off the weight
from the pan, such that the weight on the pan should now be 0.0000g.
ð
Repeat the above procedure for rest of the data entries.
90
6.10 Statistics