CD40047_R04_allPIXA_User-Manual.docx
133
Line period time
Name
Name
Line period time
Function
Sets the value for line period in microseconds to the camera.
Unit
Float
Step:
0,001
Linear
Dependency
Is only available if the
Use line period
check box has been selected.
Notes
Value must be greater than
Integration time
Camera:
Value CST
Lower limit
2k
> 16.45 µs
4k
> 29 µs
7.3k
> 47.1 µs
Upper limit
All
12.36 ms
11.2 Gain settings
Analog coarse gain
This parameter describes the CDS gain parameter of the camera which takes effect in step (1)
of analog image processing (
section 4.6.1
Analog / digital image processing
Name
Name
Analog coarse gain
Function
Pre-amplification after CCD and before analog gaining.
Channels
Red, Green, Blue
Front; Rear
Unit
-3 dB;0 dB;+3 dB; +6 dB
Log.
Dependency
None
Notes
Use this parameter for a rough adjustment of gaining at the first step.
Camera:
Value CST:
Lower limit
all
-3 dB
0 dB
3 dB
Upper limit
All
6 dB
Analog coarse gain
is a pre-amplification factor (basically to adapt the output signal of the CCD
to the input of the main analog amplifier).The output range of the CCD is greater than the input
range of the AFE (analog front-end). So with this factor the possible output range can be
adjusted. With this “adjustable two step amplification” it is possible to adjust the allPIXA for a
wide range of applications.
Line period
Line period
Line period
Int. time
Int. time
Int. time