CD40047_R04_allPIXA_User-Manual.docx
108
3
Define the area where the reference values should be calculated (a)
If it is not possible to cover the full scanning width with a white reference target
(see
sample image in
Step 4
), for example due to mechanical conditions, the resulting black
image borders on the left and right side can also be substituted by extrapolated values.
So the bright area is extended outwards.
Proceed with
Step 5
.
Specify the area from which the reference values should be calculated:
In the
Y position
box, enter the number of the first line, in the
Height
box the number of
lines to be used.
Use at least 100 lines for calculation of the reference values to eliminate noise. The
values used for the reference are calculated by averaging each column in the selected
area.
Sample image if your white reference target covers the full field of view of the camera:
Click
OK
, and then proceed with
Step 5
.