96
Chapter 3 Repair
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
4.
Calibration window, check if both coefficients CoefEtalTemp1
and 2 = 0 (
C
);
5.
If yes, close the window, if no, set them both to 0 and save chan-
ges, then close window by clicking on the cross in the top right;
■
Testing without temperature adjustment:
First, perform a comparison test on an X bowl:
1.
Access the «
Configuration
» menu, then the «
Mainte-
nance
» menu;
2.
In the synoptic view of the maintenance menu, click on "
Tes-
ting without temperature adjustment
(
A
)" (see illustration
below) + enter
1576
:
11 : Adjustments and checks for the Mixolab
Operation 11-4 Calibrating the dough temperature
Sheet : 2 /
11
Consumables:
Tools:
-
-
C
A
Содержание Mixolab
Страница 2: ......
Страница 4: ......
Страница 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Страница 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Страница 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Страница 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 39: ...Chapter 3 Repair 39 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab 2 Instructions See following pages...
Страница 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Страница 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...