4
-
This paragraph covers important warning for safety to be observed before reading the instructions.
Fully understand the following warning before reading this manual. These warnings are important for
preventing the damage to human bodies as well as accidents.
1. Switch and overcurrent protective device
This recorder is not provided with a replaceable
overcurrent protective device. Prepare a switch
and an overcurrent protective device for the power
supply (circuit breakers, circuit protectors or the
like) within 3m of this recorder in a location where
the operator can access easily Use a switch and
an overcurrent protective device conforming to
IEC947-1 and IEC947-3.
2. Be sure to ground this instrument
Before turning the power on, connect the
protective conductor terminal of this recorder to
the protective conductor of the power supply
facility. In order to prevent an accident by electric
shock, do not disconnect this connection during
operations.
3. Before turning on the power supply
In order to ensure safety, before turning on the
external power switch, make sure that the power
voltage is within the range indicated on the power
supply label.
4. Don't repair or modify this instrument
Make sure that any persons other than service
engineers approved by CHINO CORPORATION
do not repair or modify this instrument by
replacing parts. Otherwise it may be damaged or
will not function normally or an accident such as
electric shock may occur. For ordinary operation,
it is not necessary to pull out the internal unit.
5.Use this recorder following
this instruction manual
Use this recorder correctly and safely by following
this instruction manual. CHINO CORPORATION
will not be responsible for any injury, damage, lost
profit or any other claim, which may result from its
wrong use.
6. Turn off the power supply if an abnormal symptom occurs
Turn off the power supply immediately and contact your local CHINO’s sales agent if any abnormal
odor, noise or any smoke occurs, or if this recorder becomes high temperature that is too hot to be
touched.
Warning
The following fuse is mounted in the power
supply unit of this recorder for safety use.
However, this fuse is not replaceable.
Maker: Daito Communication Apparatus Co., Ltd
Model: SBL32
Fuse in the power supply
Reference
100-240V AC
50/60Hz 65VA MAX
Power supply label
Power
terminal/
Protective
conductor
terminal
Power/protective
conductor
terminal
To the protective
conductor of
power supply
facility
Power supply
Switch
Overcurrent
protective
device
(250V 3A)
L
N