143
-
K, E, J, T, R. S. B, N: IEC584, JIS C1602-1995
U (Cu-CuNi), L (Fe-CuNi): DIN43710
W-Wre26, Wre5-Wre26, PtRh40-PtRh20,
NiMo-Ni, CR-AuFe, Platinel II: ASTM
Input type
Measurement range
Reference
range
Accuracy
rating
Display
resolution
-200.0 to 300.0 °C ±13.8 mV
0.1 °C
-200.0 to
600.0 °C ±27.6 mV
0.1 °C
K
-200 to 1370 °C ±69.0 mV
1
°C
-200.0 to
200.0 °C ±13.8 mV
0.1 °C
-200.0 to
350.0 °C ±27.6 mV
0.1 °C
E
-200 to
900
°C ±69.0 mV
1
°C
-200.0 to
250.0 °C ±13.8 mV
0.1 °C
-200.0 to
500.0 °C ±27.6 mV
0.1 °C
J
-200 to
1200 °C ±69.0 mV
1
°C
-200.0 to
250.0 °C ±13.8 mV
0.1 °C
T
-200.0 to
400.0 °C ±27.6 mV
0.1 °C
0 to
1200 °C ±13.8 mV
1
°C
R
0 to
1760 °C ±27.6 mV
1
°C
0 to
1300 °C ±13.8 mV
1
°C
S
0 to
1760 °C ±27.6 mV
1
°C
B
0 to
1820 °C ±13.8 mV
±0.1
%
±1digit
1
°C
-200.0 to
400.0 °C ±13.8 mV
0.1 °C
-200.0 to
750.0 °C ±27.6 mV
0.1 °C
N
-200 to
1300 °C ±69.0 mV
1
°C
W-WRe26
0 to
2315 °C ±69.0 mV
±0.15%
±1digit
1
°C
WRe5-
Wre26
0 to
2315 °C ±69.0 mV
1
°C
PtRh40-
PtRh20
0 to
1888 °C ±13.8 mV
1
°C
-50.0 to
290.0 °C ±13.8 mV
0.1 °C
-50.0 to
600.0 °C ±27.6 mV
0.1 °C
NiMo-
Ni
-50 to
1310 °C ±69.0 mV
1
°C
CR-AuFe
0.0 to
280.0 °C ±13.8 mV
±0.2%
±1digit
0.1 °C
0.0 to
350.0 °C ±13.8 mV
0.1 °C
0.0 to
650.0 °C ±27.6 mV
0.1 °C
Platinel
Ⅱ
0 to
1395 °C ±69.0 mV
1
°C
-200.0 to
250.0 °C ±13.8 mV
0.1 °C
-200.0 to
500.0 °C ±27.6 mV
0.1 °C
U
-200.0 to
600.0 °C ±69.0 mV
±0.15%
±1digit
0.1 °C
-200.0 to
250.0 °C ±13.8 mV
0.1 °C
-200.0 to
500.0 °C ±27.6 mV
0.1 °C
Thermocouple
L
-200 to
900
°C ±69.0 mV
±0.1
%
±1digit
1
°C
Input type
Measuring range
Reference
range
Accuracy
rating
Display
resolution
-13.80 to 13.80 mV ±13.8
mV
10
μ
V
-27.60 to 27.60 mV ±27.6
mV
10
μ
V
-69.00 to 69.00 mV ±69.0 mV
10
μ
V
-200.0 to 200.0 mV ±200.0 mV
100
μ
V
-500.0 to 500.0 mV ±500.0 mV
100
μ
V
-2.000 to 2.000 V
±2 V
1 mV
-5.000 to 5.000 V
±5 V
1 mV
-10.00 to 10.00 V
±10 V
10
mV
-20.00 to 20.00 V
±20 V
10
mV
DC
Voltage
-50.00 to 50.00 V
±50 V
±0.1%
±1digit
10
mV
-140.0 to 150.0
°
C
160
Ω
±0.15%
±1digit
0.1
°
C
-200.0 to 300.0
°
C
220
Ω
0.1
°
C
Pt100
-200.0 to 850.0
°
C
400
Ω
±0.1%
±1digit
0.1
°
C
-140.0 to 150.0
°
C
160
Ω
±0.15%
±1digit
0.1
°
C
-200.0 to 300.0
°
C
220
Ω
0.1
°
C
JPt
100
-200.0 to 649.0
°
C
400
Ω
±0.1%
±1digit
0.1
°
C
Pt50
-200.0 to 649.0
°
C
220
Ω
±0.1%
±1digit
0.1
°
C
Resist
ance thermometer
Pt-Co
4.0 to 374.0 K 220
Ω
±0.15%
±1digit
0.1
K
Note) The accuracy is under the reference operation condition. For the thermocouple inputs (internal
RJ), the reference junction compensation accuracy is not included.
{
Measurement ranges, Accuracy ratings and Display resolutions
c
Exception of accuracy rating
Input range
Measuring range
Accuracy rating
K, E, J, T, L
-200 to 0
°
C
±0.2
%
±1digit
R, S
0 to 400
°
C
±0.2
%
±1digit
0 to 400
°
C
Not specified
B
400 to 800
°
C
±0.15
%
±1digit
N, U
-200 to 0
°
C
±0.3
%
±1digit
0 to 100
°
C
±4
%
±1digit
W-WRe26
100 to 400
°
C
±0.5
%
±1digit
0 to 100
°
C
±4
%
±1digit
PtRh20-PtRh5
100 to 400
°
C
±0.5
%
±1digit
0 to 300
°
C
±1.5
%
±1digit
PtRh40-PtRh20
300 to 800
°
C
±0.8
%
±1digit
0 to 20
K
±0.5
%
±1digit
CR-AuFe
20 to 50
K
±0.3
%
±1digit
Pt100(’97) 700 to 850
°
C
±0.15
%
±1digit
Pt-Co 4 to 50
K
±0.3
%
±1digit
Pt100
:
IEC751
(1995),
JIS
C1604-1997
JPt100
:
JIS
C1604-1981,
JIS
C1606-1989
Pt50
:
JIS
C1604-1981
*The indication equivalent to 1mV may vary under the test environment by EMC directives.
*Only the CE corresponding model applies.