8.TROUBLESHOOTING
24
Stimulation Modes descriptions
Mode
BURST
NORMAL
MRW SD
BI-PULSE
Pulse Frequency Fixed 100Hz
2~150Hz,
2~150Hz
2~150Hz
Channel 1 fixed at 4Hz, Channel 2 fixed at 100Hz
Pulse Width
50~300µs
50~300µs
50~300µs 50~300µs
50~300µs
Cycle Time
0.5 Sec.
Constant.
1 Sec.
6 Sec.
2 Sec.
Loading: 500
Ω
*All values have
10% tolerance.
7 ACCESSORIES
Self-Adhesive Electrodes
4 PCS.
9 V Battery
1 PC.
Lead Wires
2 PCS.
Instruction Manual
1 PC.
23
•
The LCD indicator lights up but unit
does not function properly.
•
Low Battery indicator flash.
•
None of LCD indicators light up.
1. Check all control settings. Are
they set to values prescribed by
your medical professional?
2. Are electrodes in proper position?
3. Check lead wires. Be sure all
connectors are firmly sealed.
4. Replace cord set with another to
check for broken wires.
1. Replace battery with a new one.
1. Replace battery with a new one.
If your unit does not seem to be operating correctly, refer to the chart below to determine what may be wrong.
If none of these measures correct the problem, the unit should be serviced.