Individual treatment time:
Press "INCREASE" button (triangle button) or “DECREASE” button (inverted
triangle button) to see next record of treatment time with the number of times or
previous record of treatment time with the number of times.
Press and hold Set button for 3 seconds to delete the displayed record. After the
record is deleted, the unit will make a "beeping" sound.
NOTE:
1. If the treatment time is under one minute, it will not be recorded. For example, If your
treatment time is 10 minutes and 30 seconds, the patient compliance timer will record 10
minutes, not 11 minutes.
2. The patient compliance timer can only record up to 999 minutes for each treatment.
Therefore, if you keep using the stimulator for over 999 minutes, it will only record 999
minutes and the record time will flash to mean the treatment time is over 999 minutes.
18
4.11 Turn Unit Off:
Turn both Channel Amplitude controls to OFF. Then unplug the electrode lead wires, grasping them by the plug, not
the cord. If treatment will be resumed shortly, the electrodes may be left on the skin. When the electrodes are
removed, clean the skin thoroughly with mild soap and water. If there is skin irritation, consult your prescribing
medical practitioner.
Caution: When the therapy time is completed, if the user doesn't turn off the amplitude knob, the unit will continue "beeping" every 10
seconds until the amplitude knob is turned off completely.
4.12 Patient Compliance Timer:
The patient compliance timer can memorize 60 sets of operation records; the total record time is 999 hours.
After the unit is turned off, you can start to use patient compliance timer. First, press and hold “Mode” button and
turn on the amplitude knob simultaneously to initiate patient compliance timer.