5
6
"DECREASE" control button (inverted triangle-button)
This button decreases the pulse width from 300~50µs, decreases the pulse rate from 150 to 2Hz, and decreases the
timer in continuous mode to from 90 to 5 min.
"MODE" button (round-button on the right side of the control panel)
This button selects a Stimulation-Mode. It offers the mode status from five types of stimulation modes which are
Burst, Normal, MRW (Modulated Rate and Width), SD (Strength Duration) and Bi-Pulse.
"SET" button (round-button on the middle of the control panel)
This button sets the rate, width and timer. Press the 'SET' button to enter a parameter setting mode including rate,
width and timer.
"LCD screen"
This LCD will be utilized to display stimulating mode/ pulse width/ pulse rate and to display the timer. The channel
output is indicated on the left side (Channel 1) and right side (Channel 2) of the LCD screen. The modes show on
the top of the LCD panel. The pulse rate and width show on the middle-right of the screen. The timer and clock
symbol show on the middle-left of the screen; the clock symbol will flash in final 5 min.
Battery compartment
9 Voltage battery- 1 pc
Knob Cover:
An acrylic knob cover protects Amplitude Controls from accidental user touch when
the unit is being used. After adjusting the output, remember to have the cover closed.
Lid Cover:
A panel covers the controls for Mode, Set, INCREASE and DECREASE adjustments. Your medical professional may
ask to set these controls for you and request you leave the cover in place.
Amplitude Controls:
It controls the "INTENSITY" level of stimulating pulses. These controls (located at the top of the unit) regulate the
amplitude, or intensity, of the stimulation and are the ON/OFF CONTROL. The ON indicator signal will stay lit as
long as the unit is working, and mimics the output of the electrical pulse.
Caution: If the stimulation levels are uncomfortable or become uncomfortable, reduce the stimulation
intensity to a comfortable level and contact your physician if problems persist.
"INCREASE" control button (triangle-button)
This button increases the pulse width from 50~300µs, increases the pulse rate from 2 to 150Hz, and increases the
timer from 5 to 90 mins in continuous mode.