Press "MODE" button
Press "MODE" button
Press "MODE" button
Press "MODE" button
Press "MODE" button
4.8 Adjust the Pulse Width:
The pulse width is adjustable 50~300µs in 10µs increments.
Press the SET button to enter the Pulse Width set function, then press INCREASE or DECREASE buttons to adjust
Pulse Width to the setting recommended by your medical professional.
4.9 Adjust Timer:
The timer is adjustable 5~90 minutes or continuous in 5 minute increments. Continuous option is just
the next step to 90 minutes, i.e. from 5~90 minutes to continuous and then to 5 minutes is a cycle.
During the 5 minute final count down, the clock symbol will flash once every second.
4.10 Adjust Channel Amplitude:
Turn Channel 1 or 2 clockwise. The output indication will be showed on the left side (Channel 1) and right side
(Channel 2) of the LCD screen as long as the unit is in operation. Slowly turn the Channel Amplitude control until
you reach the setting recommended by your medical professional. Repeat for the other channel, if both channels are
to be used.
Caution: If the stimulation levels are uncomfortable or become uncomfortable, reduce the stimulation intensity to a comfortable level
and contact your medical practitioner if problems persist.
16
15
Burst Mode
Normal Mode
MRW Mode
SD Mode
Bi-Pulse Mode
Adjust (1) Pulse Width (2) Timer by
"SET" & " INCREASE/DECREASE" button.
Adjust (1) Pulse Rate (2) Pulse Width (3) Timer by
"SET" & " INCREASE/DECREASE" button.
Adjust (1) Pulse Rate (2) Pulse Width (3) Timer by
"SET" & " INCREASE/DECREASE" button.
Adjust (1) Pulse Rate (2) Pulse Width (3) Timer by
"SET" & " INCREASE/DECREASE" button.
Adjust (1) Pulse Width (2) Timer by
"SET" & " INCREASE/DECREASE" button.
Operation Procedure Chart: