NOTE: Always read this instruction manual before use.
PREPARATION FOR USE
4.1 Check Battery:
Insert a fresh 9V alkaline or rechargeable battery into the battery compartment. Make sure that you are installing the
battery properly. The battery is inserted in the casing on the back of the stimulator unit. BE SURE TO MATCH
THE POSITIVE AND NEGATIVE ENDS OF THE BATTERY TO THE MARKINGS IN THE BATTERY
COMPARTMENT OF UNIT. To remove the battery cover, press and pull down following the direction of the
arrow indicated on the battery cover.
3.4 SD (Strength Duration) Mode:
Strength-Duration modulation consists of alternating modulated intensity and
pulse width, so the intensity is always increasing while the pulse width is
decreasing and vice-versa. The stimulation intensity is modulated to 62.5%
maximum of setting (width equal to setting). The pulse width is modulated to 67%
of setting (intensity equal to setting). Total cycle time is 6 seconds. Pulse rate
(from 2~150Hz) and pulse width (from 50~300µs) are fully adjustable.
3.5 Bi-Pulse Mode:
Delivers 4 pulses per second to Channel 1(i.e. the pulse rate of Channel 1 is fixed
at 4 Hz) while delivering 100 pulses per second to Channel 2 (i.e. the pulse rate of
Channel 2 is fixed at 100 Hz). Stimulation is burst on for 1.0 second, then off for 1.0
second (the illustration shows each pulse as a vertical line). Pulse width (from
50~300µs) is fully adjustable.