15 700 02 0319
Technical
Specifications
and
Conformity
30
Guidelines and manufacturer's declaration – resistance to emitted electromagnetic
interference
The Intelect RPW Lite model is intended for use in the electromagnetic environment specified below.
The customer or the user of the Intelect RPW Lite should assure that it is used in such an environment.
Immunity tests
IEC 60601
test level
Compliance level
Electromagnetic environment –
guidelines
Electrostatic
discharge (ESD) acc.
to IEC 61000-4-2
± 8 kV contact
discharge
± 15 kV air
discharge
± 8 kV contact
discharge
± 15 kV air
discharge
Floors should be wood, concrete or
ceramic tile. If the floor is covered
with synthetic material, the relative
air humidity should be at least 30%.
Electrical fast
transient disturbances
/ bursts according to
IEC 61000-4-4
± 2 kV for mains
cables
± 1 kV for input
and output cables
± 2 kV for mains
cables
± 1 kV for input
and output cables
Mains power quality should be that
of a typical commercial or hospital
environment.
Surges acc. to
IEC 61000-4-5
± 1 kV differential
mode voltage
± 2 kV common
mode voltage
± 1 kV differential
mode voltage
± 2 kV common
mode voltage
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, brief
interruptions and
fluctuations in the
supply voltage acc. to
IEC 61000-4-11
< 5% U
T
(
1
)
(> 95% dip in U
T
)
for ½ and 1 period
70% U
T
(30% dip in U
T
) for
25/30 periods
< 5% U
T
(> 95% dip in U
T
)
for 250/300 s
< 5% U
T
(
1
)
(> 95% dip in U
T
)
for ½ and 1 period
70% U
T
(30% dip in U
T
) for
25/30 periods
< 5% U
T
(> 95% dip in U
T
)
for 250/300 s
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
Intelect RPW Lite requires continued
operation during power mains
interruptions, it is recommended that
the Intelect RPW Lite be powered
from an uninterruptible power
supply.
Power frequency
(50/60 Hz) magnetic
field according to IEC
61000-4-8
30 A/m
N/A
2
The mains frequency magnetic fields
should be those of a typical business
or hospital environment.
1
NOTE U
T
is the mains alternating voltage prior to application of the test level
2
NOTE
The device contains no components that are sensitive to magnetic fields.