Cleaning, Maintenance, Overhaul
22
Contents of the overhaul kit
– 2 projectiles
– 2 guide tubes
– 2 boxes of sealing rings
– 1 O-ring guide
The overhaul kit can be ordered from your dealer.
Overhauling the handpiece
CAUTION!
Danger of injury due to pulse triggering when handpiece is
open.
• Disconnect the handpiece from the control device before
replacing the pulse transmitter.
ATTENTION!
An open-end spanner must be used to release and assemble the handpiece
shaft when overhauling the handpiece.
• Place the handpiece on a dry, clean and dust-free surface.
• Disconnect the handpiece from the control device.
•
Pull the tightly fitting guide tube out of the shaft.
If necessary, use a thin metal rod or the supplied
hexagonal spanner as a pulling tool by inserting it
through the openings in the guide tube.
• Unscrew the pulse transmitter screw cap from the
handpiece.
• Unscrew the shaft from the handpiece and pull it
out of the handpiece handle.
• Use the supplied open-end spanner for this pur-
pose.