15 700 02 0319
Principles
8
2
Principles
2.1
Physical principles
The Intelect RPW Lite is a compressed air-operated ballistic pressure wave generator.
Kinetic energy is converted into sound energy. This acoustic pulse is transmitted into
the tissue to be treated either directly or via an acoustic impedance adapter with the
help of a gel.
Physically speaking, these are radial pressure waves. The applied pressure pulse
propagates radially within the tissue and has a therapeutic effect on areas of the
tissue near the surface, in particular.
NOTE
Medical devices operating on the basis of the above principle are generally
referred to as radial shock wave systems in modern medical literature.
2.1.1
Indications
– Pain in the musculoskeletal system
– Aesthetic – non-medical – treatments to the skin, muscles, facial wrinkles,
connective tissue and tendons
2.1.2
Contraindications
Treatment with the Chattanooga Intelect RPW Lite is not permitted in the following
cases:
– Brain or spinal column in the treatment area
– Pregnancy
– Malignant tumour in the treatment area
CAUTION!
Pulses must not be applied to target areas located above air-filled
tissue nor to any regions near large nerves, vessels, the spinal col-
umn or head (except in the facial area).